Defender 5000 Indicators

6.4 Editable

If the
is set On, the submenu Link to will display. The 'Link to' contains RS232, 2
RS232,
nd
Editable
RS485, Ethernet, Wi-Fi and USB device. The default is RS232.
The output printed to the interface above will be saved as a .txt file distinguished by month. For
example, 201612.txt is the data printed to the interface during Dec. 2016.

7. LEGAL FOR TRADE

When the indicator is used in trade or a legally controlled application it must be set up, verified and
sealed in accordance with local weights and measures regulations. It is the responsibility of the
purchaser to ensure that all pertinent legal requirements are met.

7.1 Settings

Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights and measures regulations.
2. Perform a calibration as explained in Section 3.
3. Set Legal for Trade to ON in the Setup menu.
4. Without exiting the menu, turn the indicator off.
5. Disconnect power from the indicator and open the housing as explained in Section 2.3.1.
6. Set the position of the security switch SW2 to ON as shown in Figure 1-3, item 4.
7. Close the housing.
8. Reconnect power and turn the indicator on.

7.2 Verification

The local weights and measures official or authorized service agent must perform the verification
procedure.