Alarms
XII
Thopaz distinguishes between warnings, alarms and errors.
If Thopaz detects any of these situations, an acoustic warning signal sounds and a description
of the problem appears on the display. By pressing the two selection buttons simultaneously,
the acoustic alarm is suppressed for 60 seconds and the troubleshooting instructions appear
on the display.
Warning
Pressure is
maintained
✔
16
EN
Alarm
Pressure is
interrupted
✘
beep
Problem description
on display
Example:
Troubleshooting
instructions are
displayed
Internal error
The problem cannot
be rectified by the
user. The error
number
beep
is displayed.
Press simul-
taneously
Error number
beep
Acknowledge
acoustic
alarm