2. Safety Information
NOTE: Failure to comply with these Caution and Warning statements may result in injury to
personnel or equipment damage.
Do not operate the unit without sufficient solution in the tank. Do not use water
alone as the carrier bath. Water is not a satisfactory coupling agent unless a
wetting agent is used. For best results use BioSonic General Purpose, BioSonic
Germicidal Cleaner or BioSonic Enzymatic Ultrasonic Cleaning solutions.
Do not place objects directly on bottom of the tank because they may block
cavitation and cause buildup of heat in the transducer causing premature failure.
Use only BioSonic UC150 accessory baskets, cassette racks or beakers to hold
items.
Do not restrict airflow. Provide adequate ventilation by maintaining at least
2 inches clearance at the back.
Do not subject unit to shock or impact.
Do not immerse the unit in water.
Keep the display surface dry for proper functioning of device.
This product is intended for use only by trained healthcare professionals.
Use this product in accordance with the instructions in the owner's guide.
Unspecified or improper use of this product may impair safety protection.
Before cleaning items of considerable value ultrasonically, consult the item's
CAUTION!
manufacturer for cleaning precautions.
The power off selection from the keypad does not turn off main power to the
unit. Unplug the device from the power outlet to shut down the device.
This equipment may cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation measures such as
reorienting or relocating the unit or shielding the location.
Never drain tank by manually discarding solution as this may cause the unit to
stop working properly.
Use only the supplied power cord or equivalent rated for 105°C.
Do not disassemble the unit. There are no user serviceable parts inside.
Keep unit covered to reduce airborne contaminants.
Do not place fingers or hands in the ultrasonic tank while in operation.
Ensure appropriate decontamination in case of spillage of hazardous material on
or inside of the equipment through device openings.
4