Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.102. Info.CalibrationData
Info.CalibrationData.Inp1.pHas
Info.CalibrationData.Inp1.Slope
Info.CalibrationData.Inp1.Temp
Info.CalibrationData.Inp1.Date
Info.CalibrationData.Inp1.ElectrodeId
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.42.
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.103. Info.PrepData.D0.Date
Info.PrepData.D0.Time
Info.PrepData.D1.Type
Info.PrepData.D1.Date
Info.PrepData.D1.Time
identical to .D2
Data from the last correctly executed preparation.
3.2.2.104. Info.Checksums
Info.Checksums.MPList
Info.Checksums.ActualMethod
The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList:
Result of the checksum of the current measuring point list.
.ActualMethod: Result of the checksum of the current method in the working
memory. Identical methods with different method names have the
same results of the checksum.
3.2.2.105. Info.DetermData
Info.DetermData.Write
Info.DetermData.MPList.1.Attribute
Info.DetermData.ExV
Info.DetermData.MPList.1.X
Info.DetermData.MPList.1.Y
Info.DetermData.MPList.1.Z1
Info.DetermData.MPList.1.Z2
for every measuring point
Determination data in hexadecimal format. A measuring point list is available
in mode SET, KFT and MEAS.
3.2 Remote control commands
–20.00...7.00...+20.00
–9.999...1.000...+9.999
–170.0...25.0...+500.0
read only/read+write
read only/read+write
read only/read+write
read only/read+write
read only/read+write
read only/read+write
$G
read only
read only
read only
read only
read only
read only
read only
$G
read only
read only
$G
ON, OFF
117