3.2 Remote control commands
120
3.2.2.109. Info.ActualInfo.Inputs.Status
Info.ActualInfo.Inputs.Change
Info.ActualInfo.Inputs.Clear
Info.ActualInfo.Outputs.Status
Info.ActualInfo.Outputs.Change
Info.ActualInfo.Outputs.Clear
Status sends the current status of the I/O lines, Change sends the information
regarding whether a change in status of a line has taken place since the last
clearing, Clear clears the change information. For the output, there is a con-
version from binary to decimal, e.g.
| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
Line No.
| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
1
3
Output: 2
+ 2
= "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also page 173ff):
Inputs:
0
Start (pin 21)
1
Stop (pin 9)
2
Enter (pin 22)
3
Clear (pin 10)
4
Smpl Ready (pin 23)
5
pin 11
6
pin 24
7
pin 12
3.2.2.110. Info.ActualInfo.Assembly.CyclNo
$Q sends the current cycle number of the voltage measurement cycle, e.g.
"127". From the cycle number and the cycle time (see 3.2.2.118), a time
frame can be set up.
The cycle number is set to 0 on switching on the instrument, on every start
and for QuickMeas. It is incremented as long as the instrument remains
switched on.
3.2.2.111. Info.ActualInfo.Assembly.Counter.V
Info.ActualInfo.Assembly.Counter.Clear
$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the
volume counter is set to zero.
3.2.2.112. Info.ActualInfo.Assembly.Meas
$Q sends the current measured value from the assembly.
Outputs:
0
Ready (pin 5)
1
Cond. ok (pin 18)
2
Titration (pin 4)
3
EOD (pin 17)
4
Monitoring, line L4 (pin 3)
5
Error (pin 16)
6
Activate, line L6 (pin 1)
7
Pulse for recorder (pin 2)
8
not used
9
not used
10
not used
11
not used
12
not used
13
not used
read only
read only
$G
read only
read only
$G
(pin 6)
(pin 7)
(pin 8)
(pin 13)
(pin 19)
(pin 20)
read only
read only
$G
read only
758 KFD Titrino