WARNING!
When power is supplied and factory default 15 pin mating connector is
inserted, the Turbo-V 2K-G will start automatically.
The controller is furnished already mechanically and electrically connected to the
CAUTION!
pump. Detaching of the controller from the pump must be carried out by
authorized Agilent Vacuum Technologies personnel only.
When you run the Turbo-V 2K-G for the very first time, the control unit
NOTE
automatically starts the system with a special procedure which protects the
bearings from possible damages (SOFT START). The system is launched step by
step at full speed in a time variable from 10 mins up to 1 hour. After the system
has reached the full speed, the soft start procedure is disabled and following
starts are performed in the normal way.
In order to maintain the IP-54 protection level you must use only the connectors
NOTE
provided with the pump. Use connectors PN: 969-9957 or 969-9958 for the
power cable and fix it to the controller by means of the suitable retaining
bracket (see the following figure). Use this power cord and plug in conjunction
with a properly grounded power socket to avoid electrical shock and to satisfy
CE requirements.
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01 (G)
Instructions for Use
Use
293/372