Tab. 13
Output buffer (5 bytes)
Offset
Size
Unit
Byte
Byte
Res
0
1
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Range
Description
B7 – Start/Stop
0 = Stop
1 = Start
B6 – Low Speed
0 = off (set the "target speed"(In Buff byte-8) at" high
speed" (Out. Buff. byte-1))
1 = on (set the "target speed" (In Buff byte-8) at "low
speed" (Out. Buff. byte-3))
B5 – Soft Start
0 = off (this function is active at first pump run-up and can't
be deactivated by the user, after first run-up activate this
function only if the pump has been stopped for at least 3
months)
1= on
B4 – Active Stop
0 = off (after a stop command the controller stop driving the
motor, the pump will decrease its speed by friction with gas
load)
1 = on (after a stop command the controller will brake the
pump using the motor, activate this function only if you don't
use the vent valve and for few cycles at week)
B3 - Vent Valve On/Off
0 = close, 1 = open
(Valid only if in Parameter buffer Byte-15-B1 = 1, and Byte-
15-B2 = 0)
B2 – Purge Valve On/Off
0 = close, 1 = open, (Valid only if in Parameter buffer Byte-
15-B1 = 1, and Byte-15-B2 = 0)
B1 – Heather Jacket Enable.
Specific option, not available on standard models
B0 – Stop Speed Reading
0 = no (after a stop command the user can't know the real
pump speed
1 = yes (after a stop command the user can know the real
rotor speed; see In Buff. Byte Offset 6) (this function is
available for the latest controllers, otherwise will be ignored)
Technical Information
Profibus Message Mapping
(G)
17
Win
Serial
000
001
100
107
122
145
112
167
351/372