Agilent Technologies Turbo-V 2K-G Serie Manual De Instrucciones página 311

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 67
PIN N.
12
13
14
15
Signal Description
START/STOP: input signal to start or stop the pump. With the
supplied cover connector the START/STOP (+) signal is connected to
the +24 Vdc pin and the START/STOP (-) signal to the GROUND pin:
in this condition the pump doesn't start.
INTERLOCK: input signal to confirm the pump rotation. With the
supplied cover connector the INTERLOCK (+) signal is connected to
the +24 Vdc pin and the INTERLOCK (-) signal to the GROUND pin, in
this condition the pump automatically starts as soon as the controller
recognises the input supply ("Plug & Pump").
SPEED SETTING: PWM input signal to set the pump speed. The
PWM signal characteristics must be the following:
Figure 13
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
SIGNAL NAME
VENT (+)
FAULT N.O.
ANALOG OUTPUT
GROUND
frequency: 100 Hz +/-20%
amplitude: high level from 12 to 24 V
duty cycle range: from 25 % to 75 % corresponding to a pump
speed from 475 Hz to 555 Hz linearly (see the following diagram).
Technical Information
Interconnections
INPUT/OUTPUT
IN
OUT
OUT
OUT
(G)
17
ISOLATION
Optocoupled
Relay
Optocoupled
Transformer
311/372