• Disconnect the power cord
from the compressor and wind
it up.
• Clip the power cord below the
carrying handle for storage.
Store the compressor and accessories safely out of continuous
direct sunlight.
Always disconnect the compressor from the mains for storage
because electrical devices that remain plugged in present a po-
tential hazard source.
4.4 Care of the nebuliser and accessories
The nebuliser and the accessories used must be cleaned
thoroughly after each application and disinfected at least once
a week. For high risk groups (e.g., cystic fibrosis patients), the
nebuliser and the accessories used must be disinfected at
least once a day.
Cleaning
• Place all components (except
the PARI SINUS tube system)
in warm tap water and a little
dishwashing liquid for about
5 minutes. If necessary, use a
clean brush to remove loose
residue (the brush must be re-
served exclusively for this
purpose).
• Rinse all parts thoroughly in running water.
• You can remove excess water more quickly by shaking all
parts.