Hardness-Calcium
(166)
10 ml
10 ml
24 mm
24 mm
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Fill a clean cell (24mm) with 10 ml of your sample.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
2) Close the lid, place the cell into the PrimeLab and press ZERO.
4
3) Remove the cell and open the lid.
4) Add 10 drops PL Calcium Hardness 1 into vial.
5) Close the lid. Shake cell to mix liquid.
6) Add 10 drops PL Calcium Hardness 2 into vial.
7) Close the lid, place the cell into the PrimeLab and press TEST.
4
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
99
0 - 500 mg/l
Name on device: 166-Hard-Cal-liq
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(CaCO )
3
PL Calcium Hardness 1
PL Calcium Hardness 2
= continue with step 3)
= continue with step 5)
Continued...
Liquid
(POL20CH1)
(POL20CH2)