Chloride in MeOH
(167)
10 ml
24 mm
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Fill 10ml of the sample water into a clean 24mm cuvette.
2) Add 10 drops PL30CLMEOH into vial.
3) Close the lid. Shake cell to mix liquid.
4) Press OK to start 15:00 minutes countdown.
5) Swivel the vial back and forth 5 times.
6) Open the lid.
7) Clean the vial with a clean and soft cloth.
8) Place the cell into the PrimeLab do not close lid!
9) Place Turbidity-ADAPTER on top of the open cuvette.
10) Press TEST.
4
Note(s)
->
Make sure the Turbidity-ADAPTER is calibrated successfully.
->
Shake the liquid-bottle before adding the liquid to the vial.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
37
0 - 20 mg/l
Name on device: 167-Chloride-in-MeOH-liq
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
-
(Cl )
PL30CLMEOH
Liquid
37