BN LUX
(168)
10 ml
24 mm
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Fill a clean 3ml vial with 2ml of Reagent S.
2) Place 3ml vial-adapter into PrimeLab.
3) Place 3ml vial into adapter and aplpy light shield.
4) Press ZERO.
4
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
5) Remove light protection shield and 3ml vial.
6) Handle the sample according to the BN LUX procedure.
7) Press OK to start 01:00 minutes countdown.
8) Fill a clean 3ml vial with 2ml of the yellow filtrated aqueous
solution.
9) Place 3ml vial into adapter and aplpy light shield.
10) Press TEST.
4
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
21
0 - 100 mg/l
Name on device: 168-BNLUXTEST-liq
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(KOH)
TM-reagent-S
(TM149reagS)
= continue with step 2)
= continue with step 4)