Nitrogen-Total (LR)
(151)
Continued...
18) Proceed to next step to intitiate next countdown.
19) Reagents need to react press TEST to start 02:00 minutes countdown.
4
20) Open 1 Acid LR/HR vial(s) and add 2ml from ZERO vial to one.
21) Add 2ml from former TEST vial to second new vial. This is your new
22) Cap both cells and shake to mix. CAUTION: heat build-up!
23) Place the 16 mm adaptor into the device.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
!
25) Place the ZERO cell and the light protection cover and than press
4
26) Remove the cell from the device.
27) Place the sample cell and the light protection cover into the device.
4
B) Only ZERO and TEST:
1) Place the 16mm adaptor into the device.
2) Place the ZERO cell and the light protection cover and than press
4
ZERO.
3) Remove the cell from the device.
4) Place the sample cell and the light protection cover into the device.
4
Than press TEST.
169
Name on device: 151-NitroTotLR-pre
This is your new ZERO vial.
SAMPLE vial.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
3
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
ZERO.
Than press TEST.
0.5 - 25 mg/l
(N)
= continue with step 22)
= continue with step 25)
16mm vial
Continued...
169