Legionella
(147)
10 ml
24 mm
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
Please select:
2
A) Countdown + TEST
3
B) Only ZERO + TEST
4
B) Measurement procedure for B) Countdown + TEST
1) Eluate filtered particles by shaking for 2 minutes.
2) Press OK to start 02:00 minutes countdown.
4
3) Press OK to start 15:00 minutes countdown.
4
4) Press OK to start 05:00 minutes countdown.
4
5) Press OK to start 03:00 minutes countdown.
4
6) Press OK to start 10:00 minutes countdown.
4
7) Press OK to start 03:00 minutes countdown.
4
8) Press OK to start 03:00 minutes countdown.
4
9) Press OK to start 03:00 minutes countdown.
4
10) Press OK to start 03:00 minutes countdown.
4
11) Press OK to start 02:00 minutes countdown.
4
12) Press OK to start 05:00 minutes countdown.
4
13) Place filled 1ml LG-CB vial into the adapter. Place the light shield
4
(PLSp-LS-1) on top and press ZERO.
14) Place filled 1ml LG-CB vial into the adapter. Place the light shield
(PLSp-LS-1) on top and press TEST.
4
15) The determined result is immediately displayed. Press the "OK" button
4
to save the result together with date and time.
137
60 - 1000000 cfu/test
Name on device: 147-Legionella-liq
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
Legipid-Kit
(LGP-10/LGP-40/LGP-100)
2 x cuvettes for magnetic holder
Magnetic holder
(LG-MP2)
Plastic adapter for 1ml vials
Filter-kit (manual/professional)
Liquid
(LG-MCHB)
(PLSp-LegiAD-1)
(LP-Fil-man/LP-Fil-Prof)
Continued...
137