Chlorine
(12)
(free/combined/total)
Continued...
After pressing the button 4 (OK) the measurement is continued to determine the
!
total chlorine (tCl) and later also bonded chlorine (cCl). If this is not required the
measurement can be termianted by pressing "HOME". IN this case only the value
9) Remove the cuvette from the chamber and unscrew the lid.
10) Add 3 drops "PL DPD 3 C" liquid reagent to the water sample in the
cuvette.
11) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
12) After the lapse of a 2 minute countdown the total result is displayed,
divided in
2
"fCl" = "free chlorine"
3
"cCl" = "combined chlorine"
"tCl" = "total chlorine".
Scroll through the results using the buttons "arrow up" and "arrow down".
4
Press button 4 to save the result.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
-> DPD 1 A and DPD 1 B reagent MUST be added to the vial
BEFORE water sample is added to avoid mis-readings!
->
The notes under ID 11 apply here as well.
->
Liquid reagents should be stored below 10°C and above 5°C in securely
closed bottles.
42
0 - 8 mg/l
Name on device: 12-Chlorine-liq
for free chlorine (fCl) is stored.
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Cl )
2
Liquid
42