ENGLISH
PHASE 10 : DISCHARGE
When the time of the sterilization's phase ends, the valve EV2 opens (on), allowing the
pressure's discharge from the chamber up to +0.8 bar. This condition is activated for 2
minutes time (EV2 off). At the end of this phase the valve EV2 opens again (on),
allowing the complete discharge and the reaching of the environment pressure.
EV2
EV3
EV4
EV5
EV6
EV7
ON (OFF for 2 minutes) ON
OFF
OFF
ON
OFF
ON
29