4. CALIBRATIoN
ITEM
ENTER SETTING
CALIBRATIoN
WITH No LoAD
FIRST LoAD
SECoND LoAD
THIRD LoAD
ATTENTION:
1. The tare weight can't exceed the 50%F.S.The range between load1, load2, load3 can't less than 20%FS, otherwise, calibra-
tion finished automaically.
2. LOAd1 < LOAd2 < LOAd3
OPERATION
DISPLAY
USER'S GUIDE | WEIGHING HOOK | STA
SETTING RANGE & EXPLAIN
Press
together, 2 seconds.
Press
together, then input cali-
bration password.
Enter calibration.
Make sure no load and let scale stable.
Calibration ok.
Start fisrst load point calibration.
Last calibration value or default value.
Input weight, e.g. 1000kg
Hanging weight, waiting stable.
First load calibration ok.
Start second load point calibration.
Last calibration value or default value.
Input weight, e.g. 3000kg
Hanging weight, waiting stable.
Second load calibration ok.
Start third load point calibration.
Last calibration value or default value.
Input weight, e.g. 5000kg
Hanging weight, waiting stable.
Third load calibration ok.
Calibration finished, return.
ES