1. EXCITATION
2. BEFORE ITS USE
1. Use an independent electric source to prevent electronic disturbances.
2. Don't place any object on the platform when switching on the indicator.
3. Please, warm-up the scale during 2-3 minutes before using it.
4. Avoid sudden changes in temperature and draughts.
5. Don't overload the scale; do not exceed its maximum capacity.
3. CONSUMPTION
BR15
4. LOAD CELL CONNECTION TO THE INDICATOR
The connector of the load cell has 7 pins
Do not disconnect the connector of the load cell when the indicator is working, because you could damage the equipment.
•
36
BR15
Input
230~240V
Output
10V 600mA
Rechargeable
6V/4Ah
Battery
Battery life:
without back illumination, approx, 160 hours.
With back illumination, approx, 120 hours.
PIN 1
EXC +
PIN 2
SEN +
PIN 3
SIG +
PIN 4
SIG -
PIN 5
SEN -
PIN 6
EXC -
PIN 7
GND
RS232 CONNEXION
PIN 1
NC
PIN 2
RX
PIN 3
TX
PIN 4
NC
PIN 5
GND
PIN 6
NC
PIN 7
NC
PIN
8
NC
PIN
9
NC
USER'S MANUAL BR15
EN