Figure 10
pH/mV channel
pH/mV Electrode
Signal
pH/mV
1
Figure 11
Electrodes connection
Coaxial cable: To terminal 2
1
Coaxial cable: To terminal 3
2
Coaxial cable (electrode with A.T.C.): To terminals 5 or 6
3
indistinctly
Triaxial cable: To terminal 2
4
Triaxial cable: To terminal 3
5
4.5.2
Programming
Temperature
2
Use coaxial cable or triaxial cable to connect the electrode. Triaxial cable is recommended
in environment with problems caused by electromagnetic interferences. Refer to
Figure
11.
Note: For a correct pH measurement a ferrite clamp filter (included) must be added on cable on the
side of pH or redox sensor.
Note: For a correct temperature measurement a ferrite clamp (included) filter must be added con
cable on the instrument side at Pt1000 sensors.
3
1
2
The factory configuration of the instrument is:
Unit: pH.
•
Temperature: From the own channel.
•
Relays
Common
3
7
4
5
6
Triaxial cable: To terminal 4
6
Independent A.T.C.:(connected by cable code 90 44, 90 45,
etc)
Yellow and pink to terminal 5
7
Brown and white to terminal 6
8
Shield (black) to terminal 7
9
Operation
1
2
3
8
9
23