The instrument stores the measured value, the date, the hour and the position in the Data
Logger. The measurements can be viewed by date, by channel or all of them.
1. From the main menu select
2. Select
and confirm.
VIEW DATA
3. Select
CALIBRATION DATA
4. Measurement data: the instrument starts an automatic vier of the registers.
Calibration data: use de direction keys to access the calibration data of each channel.
5. Use the direction keys to navigate in the Data Logger data.
STOP key: Stops the automatic view.
UP/DOWN keys: Move forward / move back rapidly.
LEFT/RIGHT keys: First / last position.
ESC key: Quit Data Logger.
Note: The view of data starts with the last stored register.
Selecting this option MultiMeter 44 sends through the RS-232 the measured data or the
calibrations, see
5.1, Page
Erase the stored data.
Every channel of MultiMeter 44 can measure temperature.
In an instrument with 3 channels can be connected 3 temperature probes (sensors
installed in different places) or only one sensor can be used for the compensation in
different channels (sensors installed in the same measuring point).
When different temperature probes are connected, a temperature readjustment can be
done per every channel. In such way, the small differences between the probes can be
compensated or the effect of the cable length.
The allowed maximum readjustment is +1/-2 °C or +1.8/-3.6 °F.
1. In the measurement display push the OK key for 2 seconds.
Push OK and use UP/DOWN keys to select the other channel's temperature.
Note: To re-establish the factory configuration, press the "GREAT DISPLAY" key with the mark over
the readjusted temperature. Confirm with the OK key.
See Accessories and Spares at
Buffer solutions are technical buffers prepared as per DIN 19267. They are contrasted
against standard reference material (SRM) from the National Institute of Standards &
Technology, NIST, in accordance with DIN 19267. They are supplied with Certificate of
Analysis indicating data related to nominal value, uncertainty, traceability, composition,
batch number and expiration date.
The MultiMeter 44 has in memory the following data.
CONSULT DATA LOGGER
or
MEASURING DATA
57.
Section 8, Page
75.
. Confirm.