Do not look directly inside the housing during the measurement. The red light and the
•
invisible infra-red light in the pulse oximeter are harmful to your eyes.
This device is not intended for use by people (including children) with restricted
•
physical, sensory or mental skills or a lack of experience and/or a lack of knowledge,
unless they are supervised by a person who has responsibility for their safety or they
receive instructions from this person on how to use the device. Children should be
supervised around the device to ensure they do not play with it.
Neither of the displays for the pulse wave and pulse bar allows the strength of the pulse
•
or circulation to be evaluated at the measurement site. Rather, they are exclusively used
to display the current visual signal variation at the measurement site and do not enable
reliable diagnostics for the pulse.
Non-observance of the following instructions can lead to incorrect or failed measure-
ments.
There must not be any nail varnish, artificial nails or other cosmetics on the finger to
•
be measured.
Ensure that the finger nail on the finger to be measured is short enough that the
•
fingertip covers the sensor element in the housing.
Keep your hand, finger and body steady during the measurement.
•
For people with cardiac arrhythmia, the measurement values of SpO₂ and the heart
•
rate may be incorrect or the measurement may not be possible at all.
In cases of carbon monoxide poisoning, the pulse oximeter displays a measurement
•
value that is too high.
To avoid falsifying the measuring result, there should not be any strong light sources
•
(e.g. fluorescent lamps or direct sunlight) in the immediate vicinity of the pulse oxi-