The pulse
oximeter is
displaying
measurement
interruptions
or high meas-
urement value
jumps
14. Technical Data
Model no.
Measurement method
Measurement range
Accuracy
Dimensions
100006_MEDEL OXYGEN PO01_USER MANUAL_REV.00_19.04.2016.indd 36
Insufficient circulation in the
measurement finger
Measurement finger is too
large or too small
Finger, hand or body is
moving
Cardiac arrhythmia
Oxygen PO 01
Non-invasive measurement of arterial oxygen saturation of
haemoglobin and pulse rate in finger
SpO₂ 0 – 100%,
Pulse 0 – 254 beats /minute
SpO₂ 70 – 100%, ± 2%,
Pulse 30-250 bpm, ± 2 beats /minute
L 61 mm x W 36 mm x H 32 mm
Observe the warnings and safety
notes in chapter 5
Fingertip must have the following
measurements: Width between 10
and 22 mm
Thickness between 5 and 15 mm
Keep your finger, hand and body still
during the measurement.
Seek medical attention
36
10/06/2016 11:25:48