When rSl or iSl are displayed, hold the key
keys
and then press the key
LEDS AND ALARMS:
pressed unit the label on display flashes. At this point use the arrow
to confirm.
Alarm active
Hour or time
Temperature expressed in C or F.
Room probe failure. Check the connections or change the room probe.
Evaporator probe failure. Check the connections or change the evaporator probe.
Needle probe failure. If continuously is displayed, clean the probe. If the problem
continues change the probe. (if the screen shows ip3 / end, it means that the
temperature-controlled cycle is finished successfully). In time controlled cycle normally
you will see this alarm.
Real time and clock data-lost. Set the time/clock of the product. Press the
seconds. The by using SET key, set the clock as follows: Min: Minutes, Hou: Hour, dAy:
Day, Mon: Month, yEA: Year. tiM: Time format , EU or USA.
Real time clock failure. Reset the clock settings. If alarm continues call authorized
service.
Maximum temperature alarm. Check the temperature of the cabinet.
While loading to close the air blowing canals of the product. Make sure you load the
products as it is shown in the user manual.
The evaporator fan is not working.
The door is left open. Make sure you load the products as it is shown in the user manual.
Make sure that the door is closed and sealed completely.
Minimum temperature alarm. Check the temperature of the cabinet.
Digital thermostat failure. Please contact with technical service
12