Result Number
WALK AWAY MODE
REMOVE TEST DEVICE
none
WHEN PROMPTED
none
* outside of the United States contact your local BD representative
Message
Possible Cause(s)
Test device inserted at the
incorrect workflow step
INSERT
during Walk Away operation
mode.
Test device was removed
TEST DEVICE
before the test run was
WAS REMOVED
complete during Walk Away
REPEAT ASSAY
operation mode.
Corrective Action(s)
Current run aborted. Remove test device and
immediately restart the Walk Away test sequence. Insert
the test device after the display window prompt to do so.
Current run aborted. The test must be repeated with a
new test device because the elapsed development time
of the assay is unknown.
37