− Upholsteries: are assembled on a very resistant frame, with soft foam and coated with flexible laminated
PVC or leather, which besides being resistant, providing easy asepsis for the whole.
− Mechanical structure: manufactured in SAE 1020 rolled steel profiles and welded by MIG process, to
guarantee the strength and durability.
− Electrostatic painting: it is applied to all structural metal parts of the equipment. The special
polyurethane paint provides the equipment with high durability coating and has antibacterial
properties according to JIS Z 2801: 2000, where in 24 hours, the bacterial reduction is greater than
99.9%.
− Plastic covers: made of high-strength ABS with acrylic cover, the plastic covers do not require painting
and allowing polishing for recovery in case of wear or scratches.
− Electrical system: the equipment can work in frequencies of 50 or 60 Hz and can be configured to be
connected to the voltages of 118/127/220/230 Volts by an authorized technician. The maximum
supply voltage of electronic boards, motors, and other controls is 24 V. The electrical system has
On/Off switch and protection fuses.
− Spindle drive motor: all motors used in the equipment are produced by Robert Bosch of Brazil and have
advantages such as low noise, the absence of oil reservoir, uniformity in displacement, reduction of
energy consumption and low cost for maintenance. The motors also have protection systems that act
on the mechanical and electrical parts of the set.
− Pressurized suspension: the high pressure nitrogen cylinder relieves the load of the gear motor lifting
system of the equipment in 100 kg. This means that a patient with this approximate weight is not
acting on the electromechanical lifting components.
− Cabinet made of MDF (Medium Density Fiberboard): the cabinet and its drawers are made of MDF and
laminated coating, providing asepsis easiness and better finishing.
The pre-installation must be guided by the Olsen authorized service to ensure that the environment is
adequate the specified requirements for operation of the equipment (see chapter 8) and the positioning in
which it is to be used. In this step the electrical connections and, if necessary, all the water and sewage pipes,
according to the equipment's options, must be prepared.
This equipment is not designed to be installed or operated in a surgical center.
The power grid must be single-phase, has specific grounding and 10 A/3 mA DR circuit breaker. The
circuit breaker must exclusively supply for the Versatile Table and must be easy and quick to disconnect from
the mains. If the power grid presents voltage variation, the installation of a surge protection device is required.
The power switch must be turned off at the end of the workday.
There is a table below for sizing the electrical installation:
Tension (V)
118/127/220/230
This equipment should only be connected to a power source with protective grounding. There is a
risk of electric shock!
6 - General Features
7 - Installation Requirements
Wire Diameter (mm
2,5
13
)
Distance (m)
2
20
Current (A)
5,0