The Versatile Table needs special attention regarding electromagnetic compatibility and must be
installed and put into use in accordance with the electromagnetic compatibility information
presented in this chapter.
Radio frequency (RF) communication equipment, portable and mobile, can affect the Versatile
Table.
The Versatile Table is intended for use in the electromagnetic environment specified below. The customer or the user should
assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions due to voltage
fluctuations/scintillation
emissions
IEC 61000-3-3
The Versatile Table is intended for use in the electromagnetic environment specified below. The customer or the user should
assure that it is used in such an environment.
IMMUNITY Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
the power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) Magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to applications of the test level.
T
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Compliance
Grupo 1
Class B
Non-applicable
Non-applicable
Guidance and Manufacturer's Declaration - Electromagnetic Emissions - I
IEC 60601 Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
<5% U
(dip > 95% in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(dip > 95% in U
)
T
T
for 5 s
3 A/m
Electromagnetic Environment - Guidance
The Versatile Table uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference with nearby electronic equipment.
The Versatile Table is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings use for domestic purposes.
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
<5% U
(dip> 95% in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(dip> 95% in U
)
T
T
for 5 s
3 A/m
16
Electromagnetic Environment Guidance
Floors should be wood, concrete or
ceramic file. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that a
typical commercial or hospital
environment.
Mains power quality should be that a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Versatile
Table requires continued operation during
power mains interruptions, it is
recommended that the Versatile Table be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.