PULOX PO-250 Manual De Instrucciones página 29

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

This instrument has the ability to store 24 hours worth of data. It can
store the measured pulse rate and SpO
data to the computer, display the data and print reports.
a. In the main menu interface as shown in Figure 7, click the
button to select "Record", then press the button to choose
whether store the data or not, choose "on" to permit storing,
choose "off " to forbid storing.
b. If the data storage function is being turned on, when return to
the measuring interface, a flashing yellow dot would appear on
screen, which means the device is in a state of storing.
c. In the state of storing, the device is in measuring interface, the
sign "Recording" would appear on the screen in 30 seconds, and
then the screen will be automatically shut down, only a flashing
yellow dot appears on screen. If short press the button at this
moment, the sign "Recording" would appear on the screen, and
then the screen will be automatically shut down again; if long
press the button, the device would return to the measuring
interface.
d. If turning on the data storage function, the former data storage
will be automatically removed.
e. When the storage space is full, it displays "Memory is full" on
the screen, and then shut down in a few seconds. But it will still
display "Memory is full" by the next time you turn on the device
on the purpose of warning the user, and a few seconds later
enter the measuring interface.
c) Uploading the data to the PC after recording
a. Please connect the device to the computer by the data line
which is equipped with the device, then double click "SpO2
Assistant" icon which is shown as Figure 5 to open"SpO2
Assistant" program, click the 'View device stored data only' ,
and then click 'Connect'. The Software will then display "Device
Stored Data", click the'user'and click the 'Receive Data' and
click the 'OK' .
If you want to watch the data,you can click "
choose the data to open.
b. On the main menu interface, the users can upload the stored
date to computer when the symbol "Record" shows "off".
d) Exit the main menu
On the main menu interface, click the button to select "Exit", then
press the button to exit the main menu.
value accurately, transfer the
2
EN
" and then

Publicidad

Tabla de contenido
loading

Tabla de contenido