Recyclable
10. Function Specification
Information
Pulse Oxygen Saturation ( SpO2 )
Pulse Rate ( PR )
Pulse Intensity (bar-graph)
SpO2 Parameter Specification
Measuring range
Accuracy
Average value
Pulse Parameter Specification
Measuring range
Accuracy
Average pulse rate
Safety Type
Pulse Intensity
Range
Battery Requirement
1.5V (AAA size) alkaline batteries ×2
Battery working life
Two1.5 V (AAA size) 600 mAh alkaline batteries can work continually
for 24 hours
Dimensions and Weight
Dimensions
Storage and Transport
Atmospheric pressure
limitation
Display Mode
2-digit digital OLED display
3-digit digital OLED display
Bar-graph OLED display
0%~100%, (the resolution is
1%).
70%~100%:±2%, Below 70%
unspecified.
Calculate the Average value in
every 4 measure value.
The deviation between average
value and true value does not
exceed 1%.
30bpm~250bpm, (the resolution
is 1bpm)
±2bpm or±2% (select larger)
Moving calculate the Average
pulse rate every 4 cardio-beat,s
cycle.
The deviation between average
value and true value does not
exceed 1%
Interior Battery, BF Type
Continuous bar-graph display,
the higher display indicates the
stronger pulse.
58.5(L) × 31(W) × 32 (H) mm
EN