Integra Padgett B Guía De Usuario página 14

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Integra
Padgett
®
Electric Dermatome Models B and PI
Troubleshooting (continued):
A pin position guide is included with the dermatome complete kit and can be ordered as an
accessory (part #3539259). This guide is intended to determine if the oscillating pin, which
moves the blade, is properly positioned. If the oscillating pin is not properly positioned, the
blade may "chatter" or "skip," yielding unsatisfactory results. Prior to each use, the dermatome
should be checked using the pin position guide to confirm that the dermatome oscillating pin
is properly positioned.
The pin position guide should be inserted between the oscillating pin and the base of the
blade guide (as shown in the diagram). This can be done from the right. If the oscillating pin
lies anywhere within the zone clearly marked on the guide, the dermatome's oscillating pin
is properly positioned. If the pin lies on the edge or outside the zone, the dermatome should
not be used, and should be sent to the Integra Padgett Dermatome Repair Center (IPDRC)
for service. This pin position check is important because rough handling, dropping the
dermatome, or inadvertently hitting the oscillating pin may cause the oscillating pin not to be
properly positioned.
Using the Pin Position Guide (3539259)
Hold the Guide in your Right hand with the Integra logo facing
up. Lower the LEFT side of the Guide Slot over the Oscillating Pin
and against the top of the Blade Guides.
Gently slide the Pin Position Guide to the LEFT until the
Oscillating Pin touches the BOTTOM edge of the Guide Slot. If the
Oscillating Pin lies anywhere between the two vertical lines the
dermatome's oscillating pin is properly positioned for optimum
performance.
13
®
1
2

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Padgett pi

Tabla de contenido