Therapy adjustment
∙
During treatment it is possible to switch from continuous
to pulsed mode and visa versa.
∙
During treatment It is possible to increase and decrease
the set power.
Every patient has a different susceptibility to warmth and this
generally reaches a steady state only after about 5 minutes. In
between adjustment of the power is therefore necessary.
Preliminary ending of the treatment
Treatment can be ended by operation of the Stop button.
Treatment stops automatically when the treatment time has
elapsed.
9 Maintenance and troubleshooting
9.1
Cleaning
9.1.1
Cleaning of the apparatus
Switch off the unit and disconnect it from the mains supply.
The apparatus can be cleaned with a damp cloth. Use lukewarm water and a non-abrasive liquid
household cleaner (no abrasive, no alcohol content solution).
9.1.2
Cleaning of the radiators and HF-cable
The radiators can be disinfected using a cloth moistened with 70% alcohol. The radiators and HF-cable
may not be sterilized by expose to steam or gas. Clean the radiators after each use.
The radiators and HF-cable should be regularly inspected for damage and bad contacts. Damaged
radiators and a damaged HF-cable may not be used.
9.2
Fault conditions
Display's fail to light up
9.2.1
Check if the mains cable is connected to mains supply.
Check if the mains supply is available (power supply) for instance with a lamp.
o
If the lamp does not light up, check the mains fuses.
o
If the lamp lights up, check the fuses of the unit.
9.2.2
Error code Er XXX* on the display
The unit has discovered a fault using its "self-test". Switch the unit back on. If the code re-appears,
contact your supplier. The unit is probably defective.
*Er XXX: a number is mentioned on position X