4.3

Polaris 50 Mobile

The following are included in the scope of supply:
1x luminaire head with sterilizable handgrip
1x caster
1x power cable
1x lower support tube
1x upper support tube
1x cantilever arm
1x cable holder
28
1x M4 socket screw
1x A fuse FST, F 1A 250 VAC, IEC 60127-2
5 MONTAGE Polaris 50 Ceiling
bearing
The mounting must only be mounted on
ceilings having a concrete stability class B25
through B55 respectively C20/25 through
C50/60. If there is any doubt, an authorized
professional must confirm that mounting is
possible on the mounting base being used.
For other types of ceilings, a safe and
sustainable ceiling anchor is to be planned.
The transfer of the forces in the ceiling must
be ensured. The applicable regional building
regulations must be observed. A fault-free
construction is assumed.
It is essential to ensure that the ceiling
bearing or fixing elements do not come into
touch with the reinforcement components of
the solid ceiling.
Drilling must be executed by a competent
specialist.
The nuts must be tightened carefully by
means of a torque wrench.
A lightweight ceiling panel with a maximum
thickness of 20 mm can be bridged with the
fixing anchor bolt. If more than 20 mm thick,
plaster or panels are to be removed prior to
installation.
DANGER
Life-threatening danger from falling luminaire.
The actual ceiling load depends on a
variety of different factors. This can
only be accurately determined for each
particular
case
engineer.
by
a
structural