DANGER
Mounting by qualified personnel
Mounting
must be done by qualified
personnel
only.
Lack
knowledge could be life-threatening.
DANGER
Life-threatening danger from falling luminaire.
Please use the fixing anchor bolt provided.
WARNING
Risk of injury from falling parts.
Mounting must be performed by two
people.
CAUTION
Use the load data table to determine which
fastening material should be used
Take account of linkage dimension before
mounting
5.1
Load data
Polaris 50 Ceiling:
Bending moment M
B
Vertical weight F
G
5.2
Attaching the ceiling bearing
Make 4 bore hole markings
Check clearances
of
appropriate
200 Nm (147 lbf ft)
130 N (30 lbf)
Drill holes, then blow them out using bellows
Drill nominal diameter = 10 mm
Drill hole depth = 75 mm
WARNING
For safety reasons the responsible structural
engineer must be informed of any incorrect
drilling e.g. by the drilling in a reinforcement
iron, as sufficient static and load balancing may
be at risk.
CAUTION
Use protective equipment in line with tool
manufacturer's instructions.
Hold the ceiling bearing on ceiling and hit
the fixing anchor bolt with a hammer and
carefully tighten the nuts to 43 nm with a
torque wrench.
The vertical alignment of the ceiling bearing
must be considered.
6
MOUNTING Polaris 50
Ceiling
6.1 Attaching the ceiling bearing
DANGER
Mounting by qualified personnel
Mounting must be done by qualified
personnel
only.
Lack
knowledge could be life-threatening.
Please note the installation instructions for
ceiling bearing under point 5.
DANGER
Electric shocks are life-threatening.
The luminaire must be isolated from the
mains by an external all-pole switch.
EN_GB
of
appropriate
29