EN_GB
11
DISINFECTION CLEANING
AND STERILIZATION
DANGER
Electric shocks are life-threatening.
Remove the connector from the power supply
and put the switch in the "OFF" position.
CAUTION
Damage to property caused by incorrect
cleaning
No cleaning agents containing alcohol,
solvents, chlorine or abrasive products must be
used
Concentrated disinfectants may damage the
cover.
Comply with mixing ratios recommended by
manufacturers.
Unsuitable cloths can cause scratches.
NOTE
Dirt decreases luminosity
Clean regularly to keep the cover clean.
Disinfection by spraying and/or wiping is
permitted.
Clean the PMMA cover using a chamois with
glass cleaner.
Disinfectant
spraying
horizontal positions only.
NOTE
Disinfect external luminaire parts using the following
prescribed diluted products:
®
Lysoformin
®
Dismozon
®
Hexaquart
plus
®
Sagrotan
quick disinfectant cleaner
36
is
permitted
in
CAUTION
To minimize the risk of disease transmission, in
addition to complying with this user manual, you
must also comply with the applicable occupational
health and safety regulations and the requirements
of national bodies with responsibilities for hygiene
and disinfection.
11.1
Sterilisable hand grip
Sterilization
must
accordance with ISO 17665-1 (Sterilization
of health care products - Moist heat).
CAUTION
Damage to the hand grip
Do not sterilize with hot air
Package the hand grip in a sterile pouch
before sterilization.
The hand grip is designed exclusively for
damp sterilization with 3 times fractioned
pre-vacuum and saturated steam with the
following parameters:
Temperature
Overpressure
Dwell time
Vacuum drying
WARNING
Warning against burns
The hand grip must be cooled to room
temperature before use.
After sterilization check the hand grip for
mechanical integrity.
Do not use damages hand grips.
12
SAFETY INSPECTIONS
DANGER
Electric shocks are life-threatening.
Remove the connector from the power supply
and put the switch in the "OFF" position.
The connector cable must be checked for
damage at least once per year.
NOTE
Maintenance and repairs must only be carried
out by qualified electricians.
The corresponding user profile is described in
Chapter 2, Safety instructions
be
carried
out
in
134°C
2.0 bar
6 min
20 min