Problems And Solutions - NSK Surgic Pro2 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 161
Error code
Type of error
E9
Foot control error
Motor LED error
E10
(Optic motor)
E11
Control panel error
E12
Link Module error
E13
Osseo 100+ error
External terminal
E14
error
9 – 2 Problems and solutions
When a problem is detected, check the following again before requesting a repair� If none of these is applicable or if the
trouble is not remedied even after an action has been taken, a failure of this product is suspected� Contact an
authorized NSK dealer�
Problem
Plug improperly inserted to
power source
Screen is blank�
Malfunction of power switch
Power turned on while pressing
The maximum
the foot control
rotation speed and
"--" are alternately
Foot control malfunction
displayed on the
speed display�
Control unit malfunction
Cause of Error
Failure of the foot control sensor
(Hall IC)�
Self-check error�
Adhesion of the saline solution
to the motor LED�
Failure of LED circuit�
Failure of the LCD display driver�
Failure of the electrostatic
sheet�
Failure of the link cable�
Failure of the Link Module
circuit�
Failure of Bluetooth�
communication circuit in the
Link Module�
It is affected by electromagnetic
interference from electronics
Influence of electromagnetic
interference waves emitted by
electronic equipment�
Failure of Bluetooth
communication circuit in
Osseo 100+�
OS of external terminal is not
supported�
Failure of the external terminal�
Cause of Problem
Check to make sure the plug is in�
Request repair�
Remove foot from foot control�
If the problem is solved by removing the batteries, it is a foot
control malfunction� If not, it is a control unit malfunction�
Please contact an authorized NSK dealer�
Request repair�
Remove the saline solution completely� If the error
cannot be eliminated, request repair�
Request repair�
Turn off the VarioSurg 3 and re-connect the Link
Module securely� If the error cannot be eliminated,
request repair�
Keep the sufficient distance between the tip of
Osseo 100+ and the electronic equipment� If the
error cannot be eliminated, request repair�
Check the supported OS in the application� Also,
reinstall the application� If the error cannot be
eliminated, request repair�
41
Troubleshooting
Action
Action

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido