Problem
"COM ERR OS" is
Osseo 100+ not connected
displayed�
"BATTERY FC" is
Foot control battery is low
displayed�
"BATTERY OS" is
Osseo 100+ battery is low
displayed�
Cause of Problem
Press the Cancel key to cancel the error� After confirming that
nothing is blocking the signal between the control unit and
Osseo 100+, wait for a while to check if Osseo 100+ is connected
automatically� If it is not connected automatically, perform the
following procedures�
• Turn off and on both Surgic Pro2 and Osseo 100+ connected to
• Charge the battery sufficiently with the Osseo 100+ charger and
• See the Osseo 100+ manual and try the pairing again�
If this does not resolve the issue, please contact an authorized
NSK dealer�
Depress the speed control pedal of the foot control or press the
Cancel key to cancel the error�
Replace the batteries�
If this does not resolve the issue, please contact an authorized
NSK dealer�
Depress the speed control pedal of the foot control or press the
Cancel key to cancel the error�
Charge the battery sufficiently with the Osseo 100+ charger and
try the automatic connection again�
If this does not resolve the issue, please contact an authorized
NSK dealer�
Surgic Pro2�
try the automatic connection again�
43
Troubleshooting
Action