4. Heart rhythm analyzing
5. Charging procedure
6. Discharging procedure
7. Cardiopulmonary resuscitation procedure(CPR)
8. An emergency cancellation
Pads socket
Power LED
ON/OFF Button
Indicator (check pad)
Indicator (Caution)
Indicator (Analyzing)
Speacker
5.2
Assess the patient
Before using the AED7000, the operator must first check the condition of the patient. The AED7000
can be used on a patient, only if all the following conditions are present:
- Unconsciousness
- Absence of breath
- If both the conditions above exist, the AED7000 should be used immediately.
5.3
Start the AED7000 Semiautomatic External Defibrillator
The operator should first start the AED7000 to check its working condition and power condition.
5.3.1
Checking power condition
Pressing the ON/OFF button for 2~3seconds to start the AED7000, the power indicator will turn on. If
the indicator is flickering, it means that the AED7000 is in the status of low power, and the current
battery should be replaced.
5.3.2
Check the working condition
After the AED7000 starts, the AED7000 will begin to check the condition of the pads connection.
Before the pads are properly connected, the AED7000 will prompt the operator to insert the con-
nector of the pads to the socket of the AED7000 and place the pads correctly on the patient.
Check the electrodes
If the pads are improperly connected, the indicator of "Check Pad" will be flickering
Indicator (Self-testing)
Indicator (Shock)
Shock Button
19