7.6.2
Electromagnetic immunity (IEC 60601-1-2 Table 202)
The AED7000 is intended for use in the electromagnetic environment specified below. The custo-
mer or the user of the AED7000 should ensure that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic
±6kV contact
discharge (ESD)
±8kV air
IEC 61000-4-2
±2 kV for power
Electrical fast
supply lines
transient/burst
±1 kV for input/
IEC 61000-4-4
output lines
±1kV differential
Surge
mode
IEC 61000-4-5
±2kV common
mode
<5% UT(>95%
dip in UT) for 0.5
Voltage dips,
cycle
short inter-
40% UT (60% dip
ruptions and vol-
in UT) for 5 cycles
tage variations
70% UT (30%
on power supply
dip in UT) for 25
input lines
cycles
IEC 61000-4-11
<5% UT (>95 %
dip in UT) for
5 sec
Power frequency
(50/60 Hz) ma-
3 A/m
gnetic field
IEC 61000-4-8
UT is the AC mains voltage prior to application of the test level.
Compliance
Electromagnetic environment
level
guidance
Floors should be wood, concrete or
±6kV contact
ceramic tile. If floors are covered with
±8kV air
synthetic material, the relative humidity
should be at least 30 %.
Not applicable
Not applicable
Not applicable
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
3 A/m
cation in a typical commercial or hospital
environment.
31