Cleaning And Disinfecting; Troubleshooting - Mach Power HC-AED7000 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6.2

Cleaning and disinfecting

Cleaning and disinfecting of the AED7000 should be done frequently.
The followings are the guidelines of cleaning and disinfecting of the AED7000:
6.2.1
Cleaning
Before cleaning the AED7000, making sure that the AED7000 has been power off, because any clea-
ning with the machine powered on may cause a shock hazard to the cleaners.
We recommend that the cleaners use warm water to clean the AED7000. Other clean-
sers without strong solvents could also be used.
Caution!
Do not allow water or other cleaning products to enter the AED7000 unit
when cleaning the machine, as these liquids may cause damage to the
AED7000 or a shock hazard
Use a soft cloth to wipe the AED7000. Never clean the AED7000 with any type of brush.
Caution!
When cleaning the electrode pads socket of the AED7000, do not use excess
water or cleanser
6.2.2
Disinfecting
Disinfection of the AED7000 is recommended. Use only alcohol based solvents.
Caution!
Never use the following cleansers on the AED7000:
Acetone, Ammonia cleanser, Glutaraldehyde
6.3

Troubleshooting

As with any equipment, problems may occur during the operation of the AED7000. Indicators and
voice prompts will be given to the operator to indicate any identified issues with the unit.
General trouble shooting tips are noted below.
The operator could try to salve the problem which occurring prior to or during the
operation according to these trouble shootings beblow
6.3.1
The troubleshooting of attaching electrode pads
Problem description:
The voice prompt of "Check the electrode" continues to broadcast when the electrode pads