Electrical Isolation
Category
Power
External Electrical Connections
Risk Current Category
7.3

Defibrillator

Output
Energy Sequence
Charge Time from "Shock Advised"
Analysis Time
From start of analysis to shock ready
Audible Prompts
Visual Prompts
Controls
Output Energy Accuracy
Patient impedance
Waveform details
The table below provides details of the biphasic truncated exponential waveform delivered by the
AED7000 (set to 200J) when connected to resistive loads of 25 through 175 Ohms. The waveforms are
characterized by typical values for peak current (Ip), duration of the first output phase, and duration
of the second output phase. Values are within 10%.
Output Energy
(J)
Impedance (Ω)
200
Specification
Unit operates on internal battery only
No external devices are attached to the unit
Internally powered equipment with defibrillator-proof
BF type patient applied part (as per definition of IEC
60601-1 standard)
Specification
Biphasic Truncated exponential (user configurable)
150J, 150J, 200J
10 sec. to 150J
15 sec. to 200J
9 sec
Less than 20 seconds from power up
Less than 25 seconds after 6 shocks
LED prompts
LED prompts
Two buttons - ON/OFF, Shock
±15% into any impedance from 25 to 175Ω
20Ω a 200Ω
Patient
Ip1
(Amps)
25
46
50
23
75
15.3
100
11.5
125
9.2
150
7.6
175
6.5
Ip2
Phase1
Phase2
(Amps)
(ms)
(ms)
34.0
5.5
17.0
10.9
11.3
16.3
8.5
19.7
6.8
20.5
5.8
19.8
5.2
19.8
Interval
(ms)
3.3
0.7
6.6
0.7
8.9
0.7
9.6
0.7
11.2
0.7
10
0.7
10
0.7
29