Warning!
If the conditions above are observed during the inspection of the AED7000,
the AED7000 must not be used on any patient.
6.1.2

Check the battery

The AED7000 begins to check the capacity of the battery when it starts. If the battery capacity is
low, the Power LED will be flickering, otherwise the indicator will be green.
If a low battery capacity occurs, replace the current battery with a new one.
1. Open the battery compartment by removing the thumbscrews and then sliding the battery cover
up and off.
2. Locate the 3 battery contact pins in the bottom-right corner of battery compartment.
3. Push the 3-pin battery connector onto the contact pins.
4. Place the battery into the battery compartment and replace the battery cover. Slide the cover
back so that the screw holes are aligned.
5. Replace the thumbscrews.
Warning!
We recommend replacing the battery provided by the AED7000 service provi-
der or one that matches the standard battery parameters of the AED7000
Warning!
The replaced battery doesn't match the standard battery parameters may
cause serious damage to the AED7000.
6.1.3
Check the defibrillator electrodes
Inspect the packaging of any disposable DEFIBRILLATOR ELECTRODES to ensure integrity of any
seals and validity of any expiry date.
Charge the ENERGY STORAGE DEVICE once a year. Please contact our service centre or agencies to
organize this.
25