Rhythm Recognition Performance

The AED7000 algorithm exceeds the requirements of ANSI/AAMI DF39-1993.
Section 3.3.18 and the sensitivity and specificity levels recommended by the AHA Semiautomatic
External Defibrillators for Public Access Use: Recommendations for Specifying and Reporting Arrhyth-
mia Analysis Algorithm Performance.
The test database includes shockable rhythms consisting of ventricular fibrillation rhythms (>150uV)
and wide-complex ventricular tachycardia at a rate greater than 140 BPM. Non-shockable rhythms
include various sinus rhythms including supraventricular tachycardia, atrial fibrillation, atrial flutter,
sinus rhythm with PVC's, asystole, pacemaker rhythms, and ventricular tachycardia with a rate less
than 140 BPM and/or narrow complexes.
Rhythms
Shockable:
VF
Shockable:
VT
Nonshockable:
NSR
Nonshockable:
asystole
Nonshockable:
all other rhythms

8. GLOSSARY

AED
AHA
ALS
Arrhythmia
Cardiac arrest
CPR
Defibrillation
ECG
Electrocardiogram
34
Test Sample
Performance
Size
Goal
1067
>90% sensitivity
22
>75% sensitivity
>99% sensitivity
4000
(AHA)
179
>95% sensitivity
25732
>95% sensitivity
Semiautomatic external defibrillator
American Heart Association
advanced life support
Irregular rhythm of the heart muscle
Cessation of the heart muscle
Cardiopulmonary resuscitation
High-energy pulse of electricity (shock) delivered to the heart muscle
to restore normal cardiac activity
Electrocardiogram
Electrocardiogram
Conclusion
Meets the AAMI DF39 requirement
and AHA recommendation
Meets the AAMI DF39 requirement
and AHA recommendation
Meets the AAMI DF39 requirement
and AHA recommendation
Meets the AAMI DF39 requirement
and AHA recommendation
Meets the AAMI DF39 requirement