Step-by-Step Instructions
A blinking cursor and no sound signal indicates a faulty connection.
Disconnect the measurement cable from the patient and check your
connections, clean your hook connection, moisten the canal, if
necessary, and start again.
No other adjustments are necessary before starting measurement.
Connection test feature is included in ProPex
the cables:
- Connect metal part of the connection hook to the lip clip.
- "Connection test" icon
corner of the display, indicating proper connection.
- If no icon appears, the connection hook or the measurement cable
should be replaced.
Measurement cable with attached lip clip and connection hook
constitute Applied Parts of the device.
14/172
should appear in the bottom left
END-IFU-PROPEX-II-LINEAR-V08-WEB-EUR-Multilingual-2020-11
®
II in order to check