Dentsply Sirona ProPex II Manual Del Usuario página 168

Ocultar thumbs Ver también para ProPex II:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Guidance and Manufacturer's declaration - Electromagnetic Immunity
®
The ProPex
II is intended for use in the electromagnetic environment specified
below. The customer or the user of the ProPex
such an environment.
Immunity test
Electrostatic
± 8 kV contact
discharge
(ESD) IEC
± 15 kV air
61000-4-2
± 2 kV for power
Electrical fast
supply lines
transients /
bursts IEC
± 1 kV for input/
61000-4-4
output lines
± 1 kV Line-to-line
Surges IEC
61000-4-5
± 2 kV Line-to-
ground
168/172
IEC 60601-1-2
Test level
END-IFU-PROPEX-II-LINEAR-V08-WEB-EUR-Multilingual-2020-11
®
II should assure that it is used in
Compliance level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
Not applicable
± 1 kV Line-to-line
± 2 kV Line-to-
ground
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
public low-voltage power
supply network that
supplies buildings used
for domestic purposes,
commercial or hospital,
clinic environment
Mains power quality
should be that of a typical
public low-voltage power
supply network that
supplies buildings used
for domestic purposes,
commercial or hospital,
clinic environment

Publicidad

Tabla de contenido
loading

Tabla de contenido