E203
E205
E212
E213
E214
E270
E282
799 GPT Titrino
No Oven parameters: Oven not (correctly) connected.
Exit: The error message disappears on next start. If you don't wish oven
parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your
method(s).
Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
Overload in dosing element: piston is blocked.
Exit: <STOP>, <QUIT>. Rectify fault at the wet part. Mount it again onto the
dos driver and press <CLEAR>. The dosing element will be initialized.
Overload in dosing element: cock is blocked.
Exit: <STOP>, <QUIT>. Rectify fault at the wet part. Mount it again onto the
dos driver and press <CLEAR>. The dosing element will be initialized.
3.1 General rules
115