3.2. Remote control commands
152
3.2.2.46. Mode.Parameter.Evaluation.FixTime.1.Value
Parameters for STAT: Interpolation of the time at preset fraction of the final
volume. Entry as part of V(tot). The interpolated volumes are available as vari-
able C6X.
3.2.2.47. Mode.Parameter.Monitoring.MeasVal.Status
Mode.Parameter.Monitoring.MeasVal.LowLim pH:
Mode.Parameter.Monitoring.MeasVal.UpLim
Mode.Parameter.Monitoring.MeasVal.Action end, hold, wait, none
Parameters for STAT, DOC: Monitoring of limit values. Measured values (pH,
U or I) are entered in the list of measured points only if the monitoring is ac-
tive.
On limit value violation, an action can be selected:
end: Abort of the determination.
hold: Interrupt dosing until the sequence is continued by manual switching
(&m $C).
wait: Interrupt dosing until the limit value is no longer violated, then continue
automatically.
3.2.2.48. Mode.Parameter.Monitoring.Rate.Status
Mode.Parameter.Monitoring.Rate.LowLim
Mode.Parameter.Monitoring.Rate.UpLim
Mode.Parameter.Monitoring.Rate.Action
Parameters for STAT: Monitoring of limit values of the rate. Entry in ml/min.
On limit value violation, an action can be selected:
end:
Abort of the determination.
hold:
Interrupt dosing until the sequence is continued by manual
switching (& m $C).
wait:
Interrupt dosing until the limit value is no longer violated, then
continue automatically. On violation of the lower limit, the se-
quence can no longer be continued as the rate would become
even less than that when dosing was interrupted.
3.2.2.49. Mode.Parameter.Monitoring.Temp.Status
Mode.Parameter.Monitoring.Temp.LowLim
Mode.Parameter.Monitoring.Temp.UpLim
Mode.Parameter.Monitoring.Temp.Action
Parameters for STAT, DOS, DOC: Monitoring of the limit values of the tem-
perature in °C. Temperature values are entered in the list of measured points
only if the monitoring is active.
On limit value violation, an action can be selected:
end: Abort of the determination.
hold: Interrupt dosing until the sequence is continued by manual switching
(&m $C).
wait: Interrupt dosing until the limit value is no longer violated, then continue
automatically.
3.2.2.50. Mode.Parameter.Monitoring.L10Output.AssignOutput
0.01...1.00, OFF
ON, OFF
0...±20.00
U, Ipol: 0...±2000
Upol:
0...±200.0
ON, OFF
0.000...150
0.000...150
end, hold, wait, none
ON, OFF
-170.0...500.0
-170.0...500.0
end, hold, wait, none
799 GPT Titrino