3.2. Remote control commands
160
3.2.2.87. MemoryCard.FreeMem
Free memory on the card. $Q sends number of free bytes, e.g. "4928".
3.2.2.88. MemoryCard.BatteryChange
MemoryCard.BatteryChange.Date
Date for changing battery. The date will be set with "$G".
3.2.2.89. MemoryCard.List.Card.1.Name
MemoryCard.List.Card.1.Bytes
for each directory
List of all directories on the memory card with the following information:
.Name:
Name of the directory
.Bytes:
Number of bytes used by the directory
3.2.2.90. MemoryCard.List.ActDir.1.Name
MemoryCard.List.ActDir.1.Mode
MemoryCard.List.ActDir.1.Quantity
MemoryCard.List.ActDir.1.DosUnit
MemoryCard.List.ActDir.1.Bytes
MemoryCard.List.ActDir.1.Checksum
for each method
List of all methods of the current card directory with the following information:
.Name:
Name of the method
.Mode:
Mode
.Quantity: Measured quantity
.DosUnit: Buret of the method
.Bytes:
Number of bytes used by the method
.Checksum: Checksum of the method, see 3.2.2.133.
3.2.2.91. Config.Monitoring.Validation.Status
Config.Monitoring.Validation.Interval
Config.Monitoring.Validation.Counter
Config.Monitoring.Validation.ClearCount
Monitoring of validation.
.Interval:
Time interval in days for validation.
.Counter:
Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.92. Config.Monitoring.Calibration.Status
Config.Monitoring.Calibration.MeasInput
Config.Monitoring.Calibration.Interval
Config.Monitoring.Calibration.Counter
Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval:
Time interval in days for calibration from the last calibration date.
.Counter:
Time counter in days since last calibration. The counter is reset to
zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.
read only
$G
YYYY-MM-DD
read only
read only
read only
read only
read only
read only
read only
read only
ON, OFF
1...365...9999
0...9999
$G
ON, OFF
1, 2, diff
1...7...9999
0...9999
799 GPT Titrino