WEBTEC HPM7000 Manual De Instrucciones página 73

Ocultar thumbs Ver también para HPM7000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

The file size expected and maximum storage time for the measurement
are calculated and displayed according to the applicable parameters
selected.
The table below contains information on the triggers available:
Designation
Description
Keystroke
Recording of a measuring point by tapping on the
corresponding button
Level
Measurement is started on exceeding or dropping
below a limit value
Window
Measurement is started on exceeding or dropping
below one of two limit values for a defined measur-
ing range
Time
Measurement is started at a defined moment of
time
External
Measurement is started on transition of the digital
signal from High to Low
Measurement is started on transition of the digital
signal from Low to High
Channel warning
Recording of a measuring point if defined warning
value
values occur
Channel alarm
Recording of a measuring point if defined alarm
value
values occur
Fig. 34
Trigger, settings
Operation
73

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido