The rated ME device does not show basic performance features in terms of
EN60601-1, the malfunctioning of which, or rather present an unacceptable risk
to patients, operators or third parties in the event that it should present an outa-
ge or the power supply should disconnect.
Warning:
Portable RF communications equipment (radios), including accessories, such
as antenna cables and external antennas, should not be used in closer proximity
than 30 cm (12 inches) to parts and cables of the examination lamp specified
by the manufacturer. Failure to comply may result in a reduction of the device's
performance features.
Guidelines and manufacturer's declaration - electromagnetic emissions
The ri-pen instrument is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the ri-pen should ensure that
it is used in such an environment.
Emissions test
RF emissions
RF emissions
pursuant to
CISPR 11
RF emissions
RF emissions
according to
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Emissions of
voltage fluctua-
tions, flicker
IEC 61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable
Electromagnetic
environment - guidance
The ri-pen uses RF energy
exclusively for internal functions.
Therefore, its RF emissions are
very low and unlikely to interfere
with nearby electronic devices.
The ri-pen is intended for use
in all establishments, including
residential areas and those directly
connected to a public supply net-
work that also supplies buildings
used for residential purposes.
23