MAINTENANCE
Do not try to repair your Cold Rush unit as this may create a user hazard
and set basic safety of the user at risk in regard to electromagnetic
disturbances. An attempt of repairing the device will automatically void
warranty. After talking with your healthcare provider, to ensure you are
following the operating instructions properly, the unit shall be sent to
Össur for repairs or maintenance. No other company or individual is
allowed to service or repair the unit in any way.
CLEANING AND PROLONGED STORAGE
• When you have finished all of your cold therapy treatments, prior to
prolonged storage, fill the unit with one gallon of water (or enough so
the water level is at the ice fill line.
• Add 2 tablespoons of liquid bleach or 3 ounces of 3% hydrogenated
peroxide to the water and mix.
• Connect hoses to the Cold Rush Pad, turn unit on to MAX setting.
• Allow water mixture to circulate for 5 minutes.
• Turn unit off and unplug the power supply.
• Carefully drain the water mixture out of the unit and store in a cool
dark place leaving the lid slightly open. (Exposure to sunlight and
extreme heat may damage hoses and unit.)
• When reusing the unit after storage, fill the reservoir with clean water
and turn the pump on to circulate the water through the hoses and
flush the remaining water mixture out of the system (use old pad for
this procedure, discard the pad after cleaning is completed).
• Dispose water and refill the reservoir with fresh ice and water
combination to begin cold therapy treatment. Follow operating
instructions 2–10 before each use.
• Before you use the unit, inspect hoses and new pad for any tears or
breaks. Contact your healthcare provider to obtain new Cold Rush pad.
• Clean or replace insulation barrier as recommended by your
healthcare practitioner.
ELECTRICAL SPECIFICATIONS – POWER SUPPLY
Do not use any other power supply other than the one provided with the
unit.
• Model no: SA9505-120040US
• Input: 100-240V
• Input frequency: 50/60 Hz
• Output power: 4.8W
• Output Voltage: 12V
• Output Current: 0.4A
Or
• Model no: UES06WU-120050SPA
• Input: 120V US, 240V EU
• Input frequency: 60Hz US, 50Hz EU
• Output power: 6W
• Output Voltage: 12V
• Output Current: 0.5A
10

Publicidad

Tabla de contenido
loading