Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified
below. The customer or user of the device should assure that it is used in such
an environment.
Immunity
Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note 1
Note 2
*
**
IEC 60601
Test Level
3 Vrms, 150 kHz
to 80 MHz
10 V/m, 80 MHz
to 2.7 GHz
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in situations. Electromagnetic
propagation is affected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations
for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is
used exceeds the applicable RF compliance level above, the
device should be observed to verify normal operation. If
abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the device.
Over the frequency range 150 kHz to 80 MHz, field strengths
should be less than 3 V/m.
Compliance
Electromagnetic
Level
Environment – Guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the device,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
3 Vrms
d=1.2√P
10 V/m
d=1.2√P 80 MHz to
800 MHz d=2.3√P 800 MHz
to
2.7 GHz
where "P" is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey*, should be less than
the compliance level in each
frequency range**.
Interference may occur in
the vicinity of equipment
marked with the
following symbol:
14

Publicidad

Tabla de contenido
loading

Tabla de contenido